UK NEQAS ICC & ISH Ki-67 Data Reveal Differences in Performance of Primary Antibody Clones.
Ontology highlight
ABSTRACT: We examined data from 374 laboratories staining for Ki-67 as part of external quality assessment over 8 runs between 2013 and 2017 (total data sets=2601). One of 5 primary antibodies was used for 94.8% of submissions, with MIB-1 (Agilent Dako) comprising 58.8% of the total. Examining assessment score as a continuous variable showed the 30-9 (Ventana) and K2 (Leica Biosystems) clones were associated with the highest mean scores (17.0; 95% confidence interval, 16.8-17.2 and 16.3; 95% confidence interval, 15.9-16.6, respectively). Stain quality was not significantly different between them. Both were associated with significantly better staining compared with MIB-1 (Agilent Dako), MM1 (Leica Biosystems), and SP6 from various suppliers (P<0.05). Similarly, categorical assessment of "Good" versu