Evaluation of the accuracy of an electronic point-of-care analyzer to quantify blood creatinine concentration in goats.


ABSTRACT:

Background

There are no commercially available handheld blood creatinine analyzers validated in goats.

Objectives

The objective of the study was to validate the accuracy of a handheld point-of-care (POC) analyzer (Nova StatSensor) for quantifying blood creatinine concentration in goats. A secondary objective was to compare this POC against a chemistry analyzer to classify goats as normal or having mild or moderate azotemia.

Animals

Sixty-three goats admitted to a referral hospital.

Methods

Cross-sectional study. Venous blood was obtained, and creatinine concentration was measured by the POC in duplicate. Plasma was submitted for creatinine determination via the chemistry analyzer (gold standard).

Results

A total of 101 blood samples were collected from 63 goats. There was high repeatability for creatinine concentrations obtained by the POC (adjusted R2 = .97, P < .0001). Correlation of POC concentrations with those reported by the chemistry analyzer was moderate (adjusted R2 = .57, P < .0001). When correctly categorizing goats with mild azotemia, the POC demonstrated a sensitivity of 73.3% and a specificity of 88.3%. For moderate to severe azotemia, the POC had a sensitivity of 75.0% and specificity of 97.5%.

Conclusion and clinical importance

The Nova StatSensor POC provided above average accuracy for measuring blood creatinine concentration in goats compared with the gold standard test.

SUBMITTER: Boileau MJ 

PROVIDER: S-EPMC7995365 | biostudies-literature | 2021 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of the accuracy of an electronic point-of-care analyzer to quantify blood creatinine concentration in goats.

Boileau Melanie J MJ   Wagner Leslie L   Taylor Jared D JD  

Journal of veterinary internal medicine 20210224 2


<h4>Background</h4>There are no commercially available handheld blood creatinine analyzers validated in goats.<h4>Objectives</h4>The objective of the study was to validate the accuracy of a handheld point-of-care (POC) analyzer (Nova StatSensor) for quantifying blood creatinine concentration in goats. A secondary objective was to compare this POC against a chemistry analyzer to classify goats as normal or having mild or moderate azotemia.<h4>Animals</h4>Sixty-three goats admitted to a referral h  ...[more]

Similar Datasets

| S-EPMC10796056 | biostudies-literature
| S-EPMC8649338 | biostudies-literature
| S-EPMC8859550 | biostudies-literature
| S-EPMC3105183 | biostudies-literature
| S-EPMC4198086 | biostudies-literature
| S-EPMC10950697 | biostudies-literature
| S-EPMC11903296 | biostudies-literature
| S-EPMC6413365 | biostudies-literature
| S-EPMC3164578 | biostudies-literature
| S-EPMC7529235 | biostudies-literature