Unknown

Dataset Information

Development of an In Vitro Blink Model for Ophthalmic Drug Delivery.


ABSTRACT: The purpose of this study was to develop an advanced in vitro blink model that can be used to examine the release of a wide variety of components (for example, topical ophthalmic drugs, comfort-inducing agents) from soft contact lenses. The model was designed using computer-aided design software and printed using a stereolithography 3D printer. The eyelid and eyeball were synthesized from polyvinyl alcohol and silicone material, respectively. Simulated tear fluid was infused through tubing attached to the eyelid using a syringe pump. With each blink cycle, the eyelid slides and flexes across the eyeball to create an artificial tear film layer. The flow-through fluid was collected using a specialized trough. Two contact lenses, etafilcon A and senofilcon A, were incubated in 2 mL of a water

SUBMITTER: Phan CM 

PROVIDER: S-EPMC7996515 | biostudies-literature | 2021 Feb

REPOSITORIES: biostudies-literature