Simultaneous T1 , T2 , and T1ρ relaxation mapping of the lower leg muscle with MR fingerprinting.


ABSTRACT:

Purpose

To develop a novel MR-fingerprinting (MRF) pulse sequence that is insensitive to B1+ and B0 imperfections for simultaneous T1 , T2 , and T1ρ relaxation mapping.

Methods

We implemented a totally balanced spin-lock (TB-SL) module to encode T1ρ relaxation into an existing MRF framework that encoded T1 and T2 . The spin-lock module used two 180° pulses with compensatory phases to reduce T1ρ sensitivity to B1 and B0 inhomogeneities. We compared T1ρ measured using TB-SL MRF in Bloch simulations, model agar phantoms, and in vivo experiments to those with a self-compensated spin-lock preparation module (SC-SL). The TB-SL MRF repeatability was evaluated in