Fracture strength analysis of titanium insert-reinforced zirconia abutments according to the axial height of the titanium insert with an internal connection.


ABSTRACT: This study aimed to analyze fracture strength in vitro by varying the axial height of the titanium insert and the labial height of the zirconia abutment in an internal connection implant to identify the titanium insert axial height with optimal mechanical stability. Sixty implants with an internal connection system were used. Two-piece zirconia abutments were used with the titanium inserts. Combinations of different titanium insert axial heights (mm) and zirconia abutment labial heights (mm) constituted five groups: Gr1 (1-3), Gr2 (3-3), Gr3 (3-5), Gr4 (5-3), and Gr5 (5-5). After thermocycling, a fracture load test was performed with a universal testing machine. The initial deformation load and the fracture load were measured and analyzed. The fractured surface and cross-section of the spe