Effective radiation dose in radiostereometric analysis of the hip.
Ontology highlight
ABSTRACT: Background and purpose - Radiostereometric analysis (RSA) is the gold standard to study micromotion of joint replacements. RSA requires the acquisition of additional radiographs increasing the radiation dose of patients included in RSA studies. It is important to keep this dose as low as possible. Effective radiation dose (ED) measurements of RSA radiographs for different joints were done by Teeuwisse et al. some years ago using conventional radiology (CR); for total hip arthroplasty (THA), Teeuwisse et al. reported an ED of 0.150 milliSievert (mSv). With the modern digital radiography (DR) roentgen technique the ED is expected to be less.Material and methods - In this phantom study, simulating a standard patient, the ED for hip RSA radiographs is determined using DR under a variety of dif
SUBMITTER: Blom IF
PROVIDER: S-EPMC8023876 | biostudies-literature | 2020 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA