Complications and re-revisions after revisions of 528 metal-on-metal hips because of adverse reaction to metal debris.
Ontology highlight
ABSTRACT: Background and purpose - There is limited amount of evidence about optimal revision indications, technique, and implants when performing revision surgery for metal-on-metal (MoM) hip replacements due to adverse reaction to metal debris (ARMD). We assessed which factors are related to re-revisions and complications after a revision of MoM hip arthroplasty because of ARMD. We also aimed to provide information on optimal implants for these revisions.Patients and methods - 420 MoM total hip arthroplasties (THA) and 108 MoM hip resurfacings were implanted and later revised at our institution. We used Cox regression to analyze the factors associated with re-revisions and complications after a revision for ARMD.Results - A re-revision was performed on 27 THAs (6%) and 9 resurfacings (8%). The mos
SUBMITTER: Lainiala O
PROVIDER: S-EPMC8023960 | biostudies-literature | 2020 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA