ABSTRACT: Background
Patients with obstructive sleep apnea (OSA) often present with cardiovascular symptoms. Holter monitors were reported to predict sleep apnea, though were rarely used in everyday clinical practice. In this study, by comparing Holter monitoring to polysomnography (PSG), we try to find out an operable way for clinicians to use Holter to predict OSA risk.Methods
Patients (n=63) suspected of OSA underwent Holter monitoring with concurrent PSG at a sleep center. Respiration and heart rate variability (HRV) indices were calculated from the Holter and compared with PSG indices.Results
The sensitivity of the Holter-derived respiratory waveform for OSA was 90.0%, and the specificity was 82.6%. The time domain indices including standard deviation of all NN intervals
SUBMITTER: Lao M
PROVIDER: S-EPMC8024822 | biostudies-literature | 2021 Mar