A convenient method to verify the accuracy of oscillometric blood pressure monitors by the auscultatory method: A smartphone-based app.
Ontology highlight
ABSTRACT: It is recommended that oscillometric devices be calibrated by auscultation when first used, but this is difficult in practice. Here, we introduce a smartphone-based technique to verify the accuracy of blood pressure monitors (BPMs). We enrolled 99 consecutive subjects and tested 6 brands of BPMs in this study. During measurements of electronic oscillometric BPMs, Korotkoff sounds were simultaneously collected using a stethoscope head beneath a cuff connected to a smartphone, and an app named Accutension Stetho could then yield an auscultatory BP reading as a reference. Next, differences in BP between the different BPMs and Accutension Stetho were determined. The percentage of BP differences falling within 5, 10, and 15 mm Hg; the mean (MD) value; and the standard deviation (SD) of BP diffe