The performance of iPad-based noise-field perimeter versus Humphrey Field Analyser in detecting glaucomatous visual field loss.


ABSTRACT:

Background

To conduct a preliminary evaluation of the feasibility of visual field fast (VFF), a free iPad-based noise-field perimeter, in detecting glaucomatous scotomas versus the clinical-standard Humphrey visual field (HVF) test. VFF confronts subjects with a screen of flickering stimulus, allowing the immediate perception of scotomas.

Methods

This was a cross-sectional observational study of 66 glaucoma patients (66 eyes) and 30 healthy controls (30 eyes). All patients had no other visual field disorders. VFF was compared against HVF in terms of whole field and quadrants for the following: (1) correspondence in scotoma detection. (2) Agreement and correlation of the scotoma size (percentage of abnormal visual field area). (3) Test duration. Other domains tested included: