ABSTRACT: Background and purpose
Artefacts caused by dental amalgam implants present a common challenge in computed tomography (CT) and therefore treatment planning dose calculations. The goal was to perform a quantitative image quality analysis of our Artifact Management for Proton Planning (AMPP) algorithm which used gantry tilts for managing metal artefacts on Head and Neck (HN) CT scans and major vendors' commercial approaches.Materials and methods
Metal artefact reduction (MAR) algorithms were evaluated using an anthropomorphic phantom with a removable jaw for the acquisition of images with and without (baseline) metal artifacts. AMPP made use of two angled CT scans to generate one artifact-reduced image set. The MAR algorithms from four vendors were applied to the images with a