Synthetic computed tomography data allows for accurate absorbed dose calculations in a magnetic resonance imaging only workflow for head and neck radiotherapy.


ABSTRACT:

Background and purpose

Few studies on magnetic resonance imaging (MRI) only head and neck radiation treatment planning exist, and none using a generally available software. The aim of this study was to evaluate the accuracy of absorbed dose for head and neck synthetic computed tomography data (sCT) generated by a commercial convolutional neural network-based algorithm.

Materials and methods

For 44 head and neck cancer patients, sCT were generated and the geometry was validated against computed tomography data (CT). The clinical CT based treatment plan was transferred to the sCT and recalculated without re-optimization, and differences in relative absorbed dose were determined for dose-volume-histogram (DVH) parameters and the 3D volume.

Results

For overall body, the r