An experimental simulation model to assess wear of the porcine patellofemoral joint.


ABSTRACT: A range of surgical techniques and osteochondral interventions have been developed for early stage chondral/osteochondral repair interventions in the knee however, methods for functional, pre-clinical assessment of these therapies are limited. In this study, a method for simulating physiological loading and motion in the porcine patellofemoral joint was developed using a 6-axis simulator. As an example of how the method can be used, the influence of surgical positioning of osteochondral allografts in the patella on cartilage wear, deformation and damage and graft stability was investigated in this porcine patellofemoral joint model. The functional performance of allografts implanted either optimally (flush with the cartilage surface) or 1 mm proud of the cartilage surface was compared to a