Development and Validation of Measurement Traceability for In Situ Immunoassays.


ABSTRACT:

Background

Immunoassays for protein analytes measured in situ support a $2 billion laboratory testing industry that suffers from significant interlaboratory disparities, affecting patient treatment. The root cause is that immunohistochemical testing lacks the generally accepted tools for analytic standardization, including reference standards and traceable units of measure. Until now, the creation of these tools has represented an insoluble technical hurdle.

Methods

We address the need with a new concept in metrology-that is, linked traceability. Rather than calculating analyte concentration directly, which has proven too variable, we calculate concentration by measuring an attached fluorescein, traceable to NIST Standard Reference Material 1934, a fluorescein standard.

R

SUBMITTER: Torlakovic EE 

PROVIDER: S-EPMC8085580 | biostudies-literature | 2021 Apr