ABSTRACT: Purpose
To compare the biometry and prediction of postoperative refractive outcomes of four different formulae (Haigis, SRK/T, Holladay1, Barrett Universal II) obtained by swept-source optical coherence tomography (SS-OCT) biometers and partial coherence interferometry (PCI; IOLMaster ver 5.4).Methods
We compared the biometric values of SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOLMaster, Carl Zeiss Meditec, Jena, Germany). Predictive errors calculated using four different formulae (Haigis, SRKT, Holladay1, Barrett Universal II) were compared at 1 month after cataract surgery.Results
The mean preoperative axial length (AL) showed no statistically significant difference between SS-OCT and PCI (SS-OCT: 23.78 ± 0.12 mm and PCI: 23.77
SUBMITTER: Song MY
PROVIDER: S-EPMC8096100 | biostudies-literature | 2021