A detailed EIS study of boron doped diamond electrodes decorated with gold nanoparticles for high sensitivity mercury detection.


ABSTRACT: This work compares the electrochemical impedance response of polished and unpolished boron doped diamond (BDD) electrodes, during mercury detection measurements. For each substrate type both bare electrodes and electrodes decorated with average diameter 30 nm AuNPs were used, to investigate the role of AuNPs during mercury sensing with diamond electrodes. In square wave anodic stripping voltammetry (SWASV) measurements for mercury detection, the mercury ions in the electrolyte are deposited onto, then stripped from the diamond electrode surface. To investigate the different electrode performances during these steps, the EIS measurements were made at the deposition and stripping potentials, alongside scans at open circuit potential for comparison. The performance of the electrodes is assess