NeuroDAC: an open-source arbitrary biosignal waveform generator.
Ontology highlight
ABSTRACT: Objective.Researchers are developing biomedical devices with embedded closed-loop algorithms for providing advanced adaptive therapies. As these devices become more capable and algorithms become more complex, tasked with integrating and interpreting multi-channel, multi-modal electrophysiological signals, there is a need for flexible bench-top testing and prototyping. We present a methodology for leveraging off-the-shelf audio equipment to construct a biosignal waveform generator capable of streaming pre-recorded biosignals from a host computer. By re-playing known, well-characterized, but physiologically relevant real-world biosignals into a device under test, researchers can evaluate their systems without the need for expensivein vivoexperiments.Approach.An open-sour
SUBMITTER: Powell MP
PROVIDER: S-EPMC8096859 | biostudies-literature | 2021 Feb
REPOSITORIES: biostudies-literature