Early Postmarket Results with EmboTrap II Stent Retriever for Mechanical Thrombectomy: A Multicenter Experience.


ABSTRACT:

Background and purpose

EmboTrap II is a novel stent retriever with a dual-layer design and distal mesh designed for acute ischemic stroke emergent large-vessel occlusions. We present the first postmarket prospective multicenter experience with the EmboTrap II stent retriever.

Materials and methods

A prospective registry of patients treated with EmboTrap II at 7 centers following FDA approval was maintained with baseline patient characteristics, treatment details, and clinical/radiographic follow-up.

Results

Seventy patients were treated with EmboTrap II (mean age, 69.9 years; 48.6% women). Intravenous thrombolysis was given in 34.3%, and emergent large-vessel occlusions were located in the ICA (n = 18), M1 (n = 38), M2 or M3 (n = 13), and basilar