ABSTRACT: Purpose
To investigate the metal screw-in anchor failure mode and load to failure for 2 different eyelet alignments after anchor insertion in ovine humeri.Methods
Sixteen ovine humeri were dissected, and a 5-mm metal anchor with 2 nonabsorbable polyblend polyethylene sutures was inserted into them in the proximal position of the greater tuberosity. The alignment of the anchors after insertion was adjusted to make 2 test groups, each with 8 specimens: In group 1, the anchor eyelets were malpositioned, whereas in group 2, the anchor eyelets were aligned according to the manufacturer's instructions. After insertion, cyclic tests from 10 N to 180 N were performed with a frequency of 1 Hz for 200 cycles; specimens were then loaded to failure to evaluate the maximum load of the s