Factor IX assay discrepancies in the setting of liver gene therapy using a hyperfunctional variant factor IX-Padua.


ABSTRACT:

Background

Limited information exists regarding the factor IX (FIX) coagulant activity (FIX:C) measured by different assays following FIX-Padua gene therapy.

Objective

Assess for the first time FIX:C in five commonly used coagulation assays in plasma samples from hemophilia B subjects receiving FIX-Padua gene transfer.

Methods

FIX:C was compared between central (n = 1) and local laboratories (n = 5) in the study, and across four commonly used FIX:C one-stage assays and one FIX:C chromogenic assay. For comparison, samples of pooled congenital FIX-deficient plasma spiked with purified recombinant human FIX (rHFIX)-Padua protein or rHFIX (nonacog alfa) to obtain FIX:C concentrations from ~20% to ~40% were tested.

Results

FIX:C results at local laboratories strongl