Ontology highlight
ABSTRACT: Background
Several devices record and interpret patient brain activity via electroencephalogram (EEG) to aid physician assessment of anaesthetic effect. Few studies have compared EEG monitors on data from the same patient. Here, we describe a set-up to simultaneously compare the performance of three processed EEG monitors using pre-recorded EEG signals from older surgical patients.Methods
A playback system was designed to replay EEG signals into three different commercially available EEG monitors. We could then simultaneously calculate indices from the SedLine® Root (Masimo Inc., Irvine, CA, USA; patient state index [PSI]), bilateral BIS VISTA™ (Medtronic Inc., Minneapolis, MN, USA; bispectral index [BIS]), and Datex Ohmeda S/5 monitor with the Entropy™ Module (GE Healthcar
SUBMITTER: Eagleman SL
PROVIDER: S-EPMC8132885 | biostudies-literature | 2021 May
REPOSITORIES: biostudies-literature