Highly accelerated free-breathing real-time phase contrast cardiovascular MRI via complex-difference deep learning.


ABSTRACT:

Purpose

To develop and evaluate a real-time phase contrast (PC) MRI protocol via complex-difference deep learning (DL) framework.

Methods

DL used two 3D U-nets to separately filter aliasing artifact from radial real-time velocity-compensated and complex-difference images. U-nets were trained with synthetic real-time PC generated from electrocardiograph (ECG) -gated, breath-hold, segmented PC (ECG-gated segmented PC) acquired at the ascending aorta of 510 patients. In 21 patients, free-breathing, ungated real-time (acceleration rate = 28.8) and ECG-gated segmented (acceleration rate = 2) PC were prospectively acquired at the ascending aorta. Hemodynamic parameters (cardiac output [CO], stroke volume [SV], and mean velocity at peak systole [peak mean velocity]) were measured for ECG-gated segmented and DL-filtered synthetic real-time PC and compared using Bland-Altman and linear regression analyses. Additionally, hemodynamic parameters were quantified from DL-filtered, compressed-sensing (CS) -reconstructed, and gridding reconstructed prospective real-time PC and compared to ECG-gated segmented PC.

Results

Synthetic real-time PC with DL showed strong correlation (R > 0.98) and good agreement with ECG-gated segmented PC for quantified hemodynamic parameters (mean-difference: CO = -0.3 L/min, SV = -4.3 mL, peak mean velocity = -2.3 cm/s). On average, DL required 0.39 s/frame to filter prospective real-time PC, which was 4.6-fold faster than CS. Compared to CS, DL showed superior correlation, tighter limits of agreement (LOAs), better bias for peak mean velocity, and worse bias for CO and SV. Compared to gridding, DL showed similar correlation, tighter LOAs for CO and SV, similar bias for CO, and worse bias for SV and peak mean velocity.

Conclusion

The complex-difference DL framework accelerated real-time PC-MRI by nearly 28-fold, enabling rapid free-running real-time assessment of flow hemodynamics.

SUBMITTER: Haji-Valizadeh H 

PROVIDER: S-EPMC8145775 | biostudies-literature | 2021 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Highly accelerated free-breathing real-time phase contrast cardiovascular MRI via complex-difference deep learning.

Haji-Valizadeh Hassan H   Guo Rui R   Kucukseymen Selcuk S   Paskavitz Amanda A   Cai Xiaoying X   Rodriguez Jennifer J   Pierce Patrick P   Goddu Beth B   Kim Daniel D   Manning Warren W   Nezafat Reza R  

Magnetic resonance in medicine 20210315 2


<h4>Purpose</h4>To develop and evaluate a real-time phase contrast (PC) MRI protocol via complex-difference deep learning (DL) framework.<h4>Methods</h4>DL used two 3D U-nets to separately filter aliasing artifact from radial real-time velocity-compensated and complex-difference images. U-nets were trained with synthetic real-time PC generated from electrocardiograph (ECG) -gated, breath-hold, segmented PC (ECG-gated segmented PC) acquired at the ascending aorta of 510 patients. In 21 patients,  ...[more]

Similar Datasets

| S-EPMC10544487 | biostudies-literature
| S-EPMC7917162 | biostudies-literature
| S-EPMC4824680 | biostudies-literature
| S-EPMC6327532 | biostudies-literature
| S-EPMC4065644 | biostudies-literature
| S-EPMC6814292 | biostudies-literature
| S-EPMC9367083 | biostudies-literature
| S-EPMC9749305 | biostudies-literature
| S-EPMC7341067 | biostudies-literature