Evaluation of the setup discrepancy between 6D ExacTrac and cone beam computed tomography in spine stereotactic body radiation therapy.
Ontology highlight
ABSTRACT: The objective of this study was to analyze the difference in residual setup errors between 6D ExacTrac and 3D cone-beam computed tomography (CBCT) image-guided systems in spinal stereotactic body radiation therapy (SBRT). We investigated 76 patients with spinal tumors who received SBRT using Novalis Tx at our institution between January 2013 and September 2020. A Vac-lok (EZ-FIX®, Arlico Medical Company, South Korea) fixture and an assistive device, based on the region involved, were used to immobilize patients and to increase the inter-fractional setup reproducibility. The difference in the root mean square (RMS) between the 6D ExacTrac and 3D CBCT was -0.75 mm, 0.45 mm, 0.16 mm, and -0.03°; the RMS value was 1.31 mm, 1.06 mm, 0.87 mm, and 0.64°; and the standard deviation was 0.80 mm, 0.