Unknown

Dataset Information

Corneal stromal roughness after VisuMax and Intralase femtosecond laser photodisruption: An atomic force microscopy study.


ABSTRACT:

Purpose

To compare the induced corneal stromal bed roughness measured with atomic force microscopy (AFM) after LASIK flap creation with the IntraLase 60 kHz and the VisuMax femtosecond laser platforms.

Methods

Three freshly enucleated porcine eyes were operated with each femtosecond laser in this experimental study. Standard LASIK treatment parameters were used for the experiment. After LASIK flap creation, the corneal stromal roughness was assessed using a JPK NanoWizard II® AFM in contact mode immersed in liquid. Olympus OMCL-RC800PSA commercial silicon nitride cantilever tips were used. Surface measurements were made in 10 regions of the central cornea of each sample measuring 20 x 20 microns, at 512 x 512 point resolution. Roughness was measured using the root-mean-squar

SUBMITTER: Gros-Otero J 

PROVIDER: S-EPMC8158881 | biostudies-literature | 2021

REPOSITORIES: biostudies-literature