Spray coagulation with snare-tip versus argon plasma coagulation: An ex vivo study evaluating tissue effects.
Ontology highlight
ABSTRACT: Background and study aims Argon plasma coagulation (APC) is an effective and safe modality for many gastrointestinal conditions requiring hemostasis and/or ablation. However, it can be quite costly. A potentially more cost-effective alternative is snare-tip spray coagulation (SC). This study aimed to determine whether SC would be a safe and effective alternative to APC using an ex-vivo model. Methods Using two resected porcine stomach, 36 randomized gastric areas were ablated for 2 seconds with either APC at 1.0 L/min 20 W (APC20) and 1.4 L/min 40 W (APC40) or SC with Effect 2 60 W (SC60) and 80 W (SC80) from 3 mm. Extent of tissue injury was then analyzed histopathologically. Results The mean coagulation depth was 790 ± 159 µm and 825 ± 467 µm for SC60 (n = 9) and