Enhancing the Long-Term Stability of a Polymer Dot Glucose Transducer by Using an Enzymatic Cascade Reaction System.


ABSTRACT: Impaired glucose metabolism in diabetes causes severe acute and long-term complications, making real-time detection of blood glucose indispensable for diabetic patients. Existing continuous glucose monitoring systems are unsuitable for long-term clinical glycemic management due to poor long-term stability. Polymer dot (Pdot) glucose transducers are implantable optical nanosensors that exhibit excellent brightness, sensitivity, selectivity, and biocompatibility. Here, it is shown that hydrogen peroxide-a product of glucose oxidation in Pdot glucose sensors-degrades sensor performance via photobleaching, reduces glucose oxidase activity, and generates cytotoxicity. By adding catalase to a glucose oxidase-based Pdot sensor to create an enzymatic cascade, the hydrogen peroxide product of gluco