Delivery sheath tip invagination - An unusual cause of failure to retrieve duct occluder.


ABSTRACT: Transcatheter closure of patent ductus arteriosus is the standard of care. Retrieval of a duct occluder device is generally easy until it is detached from the delivery cable. We report two instances of failed retrieval of the device due to sheath tip invagination. The report highlights the importance of prompt identification of the mechanism of unforeseen complications in managing them effectively.

SUBMITTER: Gupta A 

PROVIDER: S-EPMC8174637 | biostudies-literature | 2021 Apr-Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Delivery sheath tip invagination - An unusual cause of failure to retrieve duct occluder.

Gupta Anunay A   Gupta Saurabh Kumar SK   Ramakrishnan Sivasubramanian S   Kothari Shyam S SS  

Annals of pediatric cardiology 20210410 2


Transcatheter closure of patent ductus arteriosus is the standard of care. Retrieval of a duct occluder device is generally easy until it is detached from the delivery cable. We report two instances of failed retrieval of the device due to sheath tip invagination. The report highlights the importance of prompt identification of the mechanism of unforeseen complications in managing them effectively. ...[more]