Unknown

Dataset Information

Field validation of a magneto-optical detection device (Gazelle) for portable point-of-care Plasmodium vivax diagnosis.


ABSTRACT: A major challenge for malaria is the lack of tools for accurate and timely diagnosis in the field which are critical for case management and surveillance. Microscopy along with rapid diagnostic tests are the current mainstay for malaria diagnosis in most endemic regions. However, these methods present several limitations. This study assessed the accuracy of Gazelle, a novel rapid malaria diagnostic device, from samples collected from the Peruvian Amazon between 2019 and 2020. Diagnostic accuracy was compared against microscopy and two rapid diagnostic tests (SD Bioline and BinaxNOW) using 18ssr nested-PCR as reference test. In addition, a real-time PCR assay (PET-PCR) was used for parasite quantification. Out of 217 febrile patients enrolled and tested, 180 specimens (85 P. vivax and 95 ne

SUBMITTER: Valdivia HO 

PROVIDER: S-EPMC8219132 | biostudies-literature | 2021

REPOSITORIES: biostudies-literature