Performance Characteristics of Long Axial Field-of-View PET Scanners with Axial Gaps.
Ontology highlight
ABSTRACT: The introduction of long (>60 cm) axial field-of-view (LAFOV) PET systems has shown their potential for clinical and research applications. LAFOV scanners are expensive, so there is interest in designing systems with longer axial coverage while mitigating cost by introducing detector gaps. We used measurements on the PennPET Explorer (64-cm AFOV prototype) and simulations of scanners up to 143-cm long to assess scanner performance with axial gaps introduced by varying the number of detector rows in each ring. Removing detectors reduces the total sensitivity and results in a non-uniform axial noise profile. Axial resolution shows small (<0.5 mm) loss from the edge of the AFOV to the center, even for a 143-cm AFOV scanner with an unrestricted acceptance angle. The presence of large axial gap
SUBMITTER: Daube-Witherspoon ME
PROVIDER: S-EPMC8224406 | biostudies-literature | 2021 May
REPOSITORIES: biostudies-literature
ACCESS DATA