Unknown

Dataset Information

Direct Mercury Detection in Landfill Leachate Using a Novel AuNP-Biopolymer Carbon Screen-Printed Electrode Sensor.


ABSTRACT: A novel Au nanoparticle (AuNP)-biopolymer coated carbon screen-printed electrode (SPE) sensor was developed through the co-electrodeposition of Au and chitosan for mercury (Hg) ion detection. This new sensor showed successful Hg2+ detection in landfill leachate using square wave anodic stripping voltammetry (SWASV) with an optimized condition: a deposition potential of -0.6 V, deposition time of 200 s, amplitude of 25 mV, frequency of 60 Hz, and square wave step voltage of 4 mV. A noticeable peak was observed at +0.58 V associated with the stripping current of the Hg ion. The sensor exhibited a good sensitivity of ~0.09 μA/μg (~0.02 μA/nM) and a linear response over the concentration range of 10 to 100 ppb (50-500 nM). The limit of detection (LOD) was 1.69 ppb, which is significantly lower

SUBMITTER: Hwang JH 

PROVIDER: S-EPMC8229311 | biostudies-literature | 2021 Jun

REPOSITORIES: biostudies-literature