A Randomized Comparison of Delivered Energy in Cardioversion of Atrial Fibrillation: Biphasic Truncated Exponential Versus Pulsed Biphasic Waveforms.
Ontology highlight
ABSTRACT: A few randomized trials have compared impedance-compensated biphasic defibrillators in clinical use. We aim to compare pulsed biphasic (PB) and biphasic truncated exponential (BTE) waveforms in a non-inferiority cardioversion (CVS) study. This was a prospective monocentric randomized clinical trial. Eligible patients admitted for elective CVS of atrial fibrillation (AF) between February 2019 and March 2020 were alternately randomized to treatment with either a PB defibrillator (DEFIGARD TOUCH7, Schiller Médical, Wissembourg, France) or a BTE high-energy (BTE-HE) defibrillator (LIFEPAK15, Physio-Control Inc., Redmond, WA, USA). Fixed-energy protocol (200-200-200 J) was administered. CVS success was accepted if sinus rhythm was restored at 1 min post-shock. The study design considered non-in