Unknown

Dataset Information

0

Shot-noise limited, supercontinuum-based optical coherence tomography.


ABSTRACT: We present the first demonstration of shot-noise limited supercontinuum-based spectral domain optical coherence tomography (SD-OCT) with an axial resolution of 5.9 μm at a center wavelength of 1370 nm. Current supercontinuum-based SD-OCT systems cannot be operated in the shot-noise limited detection regime because of severe pulse-to-pulse relative intensity noise of the supercontinuum source. To overcome this disadvantage, we have developed a low-noise supercontinuum source based on an all-normal dispersion (ANDi) fiber, pumped by a femtosecond laser. The noise performance of our 90 MHz ANDi fiber-based supercontinuum source is compared to that of two commercial sources operating at 80 and 320 MHz repetition rate. We show that the low-noise of the ANDi fiber-based supercontinuum source improves the OCT images significantly in terms of both higher contrast, better sensitivity, and improved penetration. From SD-OCT imaging of skin, retina, and multilayer stacks we conclude that supercontinuum-based SD-OCT can enter the domain of shot-noise limited detection.

SUBMITTER: Rao D S S 

PROVIDER: S-EPMC8239030 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC8448444 | biostudies-literature
| S-EPMC3018058 | biostudies-other
| S-EPMC8211087 | biostudies-literature
| S-EPMC7813855 | biostudies-literature
| S-EPMC5116674 | biostudies-literature
| S-EPMC5363080 | biostudies-literature