Unknown

Dataset Information

Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma.


ABSTRACT:

Background

This study investigated the agreement between a new rebound tonometer, IC200, and IcarePRO and Goldmann applanation tonometry (GAT).

Methods

This was a prospective cross-sectional study. We measured the intraocular pressure (IOP) in 145 eyes of 145 glaucoma patients in the sitting position using GAT, IcarePRO, and IC200. IcarePRO and IC200 measurements were also obtained in the supine position. IC200 measurement was performed using two modes: single six (IC200-single) and automatic (IC200-continuous) six-measurements mode.

Results

All tonometers provided high reproducibility in both positions (all intraclass correlation coefficients > 0.90), although it was highest with GAT, followed by IC200-continuous and IC200-single and then IcarePRO. In the sitting pos

SUBMITTER: Nakakura S 

PROVIDER: S-EPMC8247177 | biostudies-literature | 2021 Jul

REPOSITORIES: biostudies-literature