HbA1c and Glucose Management Indicator Discordance: A Real-World Analysis.


ABSTRACT: Background: There can be marked discordance between laboratory and estimated (using the glucose management indicator [GMI]) glycated hemoglobin (HbA1c) from continuous glucose monitoring (CGM). This may cause errors in diabetes management. This study evaluates discordance between laboratory and CGM-estimated HbA1c (eA1C). Methods: We performed a retrospective review of patients with diabetes who use CGM. The patients were seen at the University of Washington (UW) Diabetes Care Center from 2012 to 2019. We used UW's Institute of Translational Health Sciences to extract eligible encounters from the electronic medical record. We required that patients use CGM and that HbA1c and sensor data be obtained fewer than 4 weeks apart. There were no exclusion criteria. We c