ABSTRACT: Background
Cardiac resynchronization therapy (CRT) response assessment relies on the QRS complex narrowing criterion. Yet one third of patients do not improve despite narrowed QRS after implantation. Electromechanical wave imaging (EWI) is a quantitative echocardiography-based technique capable of noninvasively mapping cardiac electromechanical activation in three dimensions. The aim of this exploratory study was to investigate the EWI technique, sensitive to ventricular dyssynchrony, for informing CRT response on the day of implantation.Methods
Forty-four patients with heart failure with left bundle branch block or right ventricular (RV) paced rhythm and decreased left ventricular ejection fraction (LVEF; mean, 25.3 ± 9.6%) underwent EWI without and with CRT within 24 hour