ABSTRACT: Background
The goal of this work was to determine the quantitative accuracy and optimal reconstruction parameters for 124I-PET imaging in the presence of therapeutic levels of 131I. In this effort, images were acquired on a GE D710 PET/CT scanner using a NEMA IEC phantom with spheres containing 124I and increasing amounts of 131I activity in the background. At each activity level, two scans were acquired, one with the phantom centered in the field of view (FOV) and one 11.2 cm off-center. Reconstructions used an ordered subset expectation maximization algorithm with up to 100 iterations of 16 subsets, with and without time-of-flight (TOF) information. Results were evaluated visually and by comparing the 124I activity relative t