Compact and adjustable compensator for AOD spatial and temporal dispersion using off-the-shelf components.
Ontology highlight
ABSTRACT: Random access multiphoton microscopy using two orthogonal acousto-optic deflectors (AODs) allows sampling only particular regions of interest within a plane, greatly speeding up the sampling rate. AODs introduce spatial and temporal dispersions, which distort the point spread function and decrease the peak intensity of the pulse. Both of these effects can be compensated for with a single dispersive element placed a distance before the AODs. An additional acousto-optic modulator, a custom cut prism, and a standard prism used with additional cylindrical optics have been demonstrated. All of these introduce additional cost or complexity and require an extended path length to achieve the needed negative group delay dispersion (GDD). By introducing a telescope between a transmission grating and
SUBMITTER: Yamaguchi A
PROVIDER: S-EPMC8281507 | biostudies-literature | 2021 Apr
REPOSITORIES: biostudies-literature