Unknown

Dataset Information

Comparative study of renal drainage with different ureteral stents subject to extrinsic ureteral obstruction using an in vitro ureter-stent model.


ABSTRACT:

Background

To compare the efficacy of different ureteral stents subject to extrinsic ureteral obstruction (EUO), in a controlled in vitro stented ureter experiment.

Methods

We employ an in vitro ureter-stent experimental set-up, with latex tubing simulating flexible ureters attached to vessels simulating renal units and bladders. The flow behavior of five ureteral stents-polymeric 8F, tandem 6F, tandem 7F, endopyelotomy and metal-was tested under a ureteral deformation configuration of 40°, with 2000 g external force over a 3.5 cm length of the ureter. A constant fluid flow was applied through the ureter-stent configurations, and pressure fluctuations in the renal unit were monitored. We considered a renal unit pressure of 10 cmH2O or flow discontinuation in the b

SUBMITTER: Shilo Y 

PROVIDER: S-EPMC8281631 | biostudies-literature | 2021 Jul

REPOSITORIES: biostudies-literature