ABSTRACT: Purpose
The purpose is: (a) Relate magnetic resonance imaging (MRI) quality recommendations for radiation therapy (RT) to B0 field homogeneity; (b) Evaluate manufacturer specifications of B0 homogeneity for 34 commercial whole-body MRI systems based on the MRI quality recommendations and RT application; (c) Measure field homogeneity in five commercial MRI systems and one commercial MRI-Linac used in RT and compare the results with their B0 homogeneity specifications.Methods
Magnetic resonance imaging quality recommendations for spatial integrity, image blurring, fat saturation, and null banding in RT were developed based on the literature. Guaranteed (maximum) and typical B0 field homogeneity specifications for various diameter sp