ABSTRACT: Purpose
To assess the agreement and repeatability of horizontal visible iris diameter (HVID) or white-to-white (WTW) measurements between four imaging modalities; combination slit scanning elevation/Placido tomography, infrared biometry, dual rotating scheimpflug camera/Placido tomography, and swept source anterior segment optical coherence tomography (AS-OCT).Methods
A prospective study of 35 right eyes of healthy volunteers were evaluated using the Orbscan IIz, IOL Master 700, Galilei G2, and DRI Triton OCT devices. The inter-device agreement and repeatability of HVID/WTW measurements for each device were analysed.Results
Mean HVID/WTW values obtained by the Orbscan IIz, IOL Master 700, Galilei G2 and DRI Triton OCT were 11.77 ± 0.40 mm, 12.40 ± 0.43 mm, 12.25 ± 0