Non-invasive hemoglobin measurement devices require refinement to match diagnostic performance with their high level of usability and acceptability.
Ontology highlight
ABSTRACT: Anemia remains an important global health problem. Inexpensive, accurate, and noninvasive solutions are needed to monitor and evaluate anemia in resource-limited settings. We evaluated the performance of multiple point-of-care hemoglobin devices, including a novel noninvasive smartphone application tested on Apple® and Android® cell phones, Masimo Pronto®, and HemoCue® Hb-301 and Hb-801, against a gold-standard hematology analyzer (reference hemoglobin) using venous blood. We examined correlations between hemoglobin devices and reference hemoglobin, device accuracy (average bias, Bland-Altman plots, clinical performance) and classification bias (sensitivity, specificity) among 299 refugees (10mo-65y) in Atlanta, GA. Semi-structured interviews (n = 19) with participants and staff assessed u