Comparison of structural MRI brain measures between 1.5 and 3 T: Data from the Lothian Birth Cohort 1936.


ABSTRACT: Multi-scanner MRI studies are reliant on understanding the apparent differences in imaging measures between different scanners. We provide a comprehensive analysis of T1 -weighted and diffusion MRI (dMRI) structural brain measures between a 1.5 T GE Signa Horizon HDx and a 3 T Siemens Magnetom Prisma using 91 community-dwelling older participants (aged 82 years). Although we found considerable differences in absolute measurements (global tissue volumes were measured as ~6-11% higher and fractional anisotropy [FA] was 33% higher at 3 T than at 1.5 T), between-scanner consistency was good to excellent for global volumetric and dMRI measures (intraclass correlation coefficient [ICC] range: .612-.993) and fair to good for 68 cortical regions (FreeSurfer) and cortical surface measure