Laryngeal mask airway protector generates higher oropharyngeal leak pressures compared to the laryngeal mask airway supreme: A randomized clinical trial in the ambulatory surgery unit.


ABSTRACT:

Background and aims

The Laryngeal Mask Airway (LMA) Protector™ is one of the latest introduced supraglottic airway devices. It provides access and functional separation of the respiratory and digestive tracts. Compared to the LMA Supreme™, it has two digestive ports, one to provide suction in the pharyngeal region and one for gastric tube insertion. High oropharyngeal leak pressure is a marker for safe ventilation when using LMA devices. We hypothesized that oropharyngeal leak pressure of the LMA Protector™ is 5 cm H2O higher than the oropharyngeal leak pressure of the LMA Supreme™ at various cuff volumes. Secondary outcome measures were ease of insertion of both masks, fiberoptic confirmation of correct positioning, failures of insertion, presence of blood staining, sor