ABSTRACT: Purpose
The purpose of this study was to measure gantry angle-related eddy currents in a 0.35-T MRI-Linac and determine if B0 (zeroth order) eddy currents are the primary cause of gantry angle-dependent imaging isocenter shifts vs other potential causes like B0 inhomogeneities and gradient (first order) eddy currents. For conventional Cartesian acquisitions, B0 eddy currents can cause imaging isocenter shifts along both phase encode and readout directions. Gradient eddy currents can cause spatial distortion along both the phase encode and readout directions. Center frequency offsets can cause imaging isocenter shifts along the readout direction that vary with readout gradient polarity.Methods
MRI-related eddy currents and imaging isocenter s