The chylous turbidity of blood samples is one of the causes of false-high hemoglobin (Hgb) concentration measurements by the colorimetric method, which has been widely applied in hematology analyzers. In such cases, additional manual procedures are required to correct Hgb concentrations. We therefore examined the effectiveness of an optical method for measuring Hgb concentrations in samples with chylous turbidity using Hgb-O in the reticulocyte channel equipped in XN-series analyzers (Sysmex, Ko ...[more]