An MRI-compatible varus-valgus loading device for whole-knee joint functionality assessment based on compartmental compression: a proof-of-concept study.


ABSTRACT:

Objective

Beyond static assessment, functional techniques are increasingly applied in magnetic resonance imaging (MRI) studies. Stress MRI techniques bring together MRI and mechanical loading to study knee joint and tissue functionality, yet prototypical axial compressive loading devices are bulky and complex to operate. This study aimed to design and validate an MRI-compatible pressure-controlled varus-valgus loading device that applies loading along the joint line.

Methods

Following the device's thorough validation, we demonstrated proof of concept by subjecting a structurally intact human cadaveric knee joint to serial imaging in unloaded and loaded configurations, i.e. to varus and valgus loading at 7.5 kPa (= 73.5 N), 15 kPa (= 147.1 N), and 22.5 kPa (= 220.6 N). Follow

SUBMITTER: Said O 

PROVIDER: S-EPMC8302563 | biostudies-literature | 2020 Dec

REPOSITORIES: biostudies-literature