ABSTRACT: Study objectives
To assess the diagnostic performance of a nonintrusive device placed under the mattress to detect sleep apnea syndrome.Methods
One hundred eighteen patients suspected to have obstructive sleep apnea syndrome completed a night at a sleep clinic with a simultaneous polysomnography (PSG) and recording with the Withings Sleep Analyzers. PSG nights were scored twice: first as simple polygraphy, then as PSG.Results
Average (standard deviation) apnea-hypopnea index from PSG was 31.2 events/h (25.0) and 32.8 events/h (29.9) according to the Withings Sleep Analyzers. The mean absolute error was 9.5 events/h. The sensitivity, specificity, and area under the receiver operating characteristic curve at thresholds of apnea-hypopnea index ≥ 15 events/h were, respe