Unknown

Dataset Information

Reliability of mechanical ventilation during continuous chest compressions: a crossover study of transport ventilators in a human cadaver model of CPR.


ABSTRACT:

Background

Previous studies have stated that hyperventilation often occurs in cardiopulmonary resuscitation (CPR) mainly due to excessive ventilation frequencies, especially when a manual valve bag is used. Transport ventilators may provide mandatory ventilation with predetermined tidal volumes and without the risk of hyperventilation. Nonetheless, interactions between chest compressions and ventilations are likely to occur. We investigated whether transport ventilators can provide adequate alveolar ventilation during continuous chest compression in adult CPR.

Methods

A three-period crossover study with three common transport ventilators in a cadaver model of CPR was carried out. The three ventilators 'MEDUMAT Standard²', 'Oxylog 3000 plus', and 'Monnal T60' represent three

SUBMITTER: Orlob S 

PROVIDER: S-EPMC8316711 | biostudies-literature | 2021 Jul

REPOSITORIES: biostudies-literature