Computational Fluid Dynamics of Intraocular Silicone Oil Tamponade.


ABSTRACT:

Purpose

To investigate the behavior of silicone oil (SiO) at the steady equilibrium and during saccades and calculate SiO-retina contact, shear stress (SS), and shear rate (SR).

Methods

A 24 mm phakic eye mesh model underwent 50°/0.137s saccade. The vitreous chamber compartment was divided into superior and inferior 180° sectors: lens, pre-equator, postequator, and macula. SiO-retina contact was evaluated as a function of fill percentages between 80% and 90% for a standing patient, 45° upward gaze, and supine. SS and SR for 1000 mPa-s (SiO1000) and 5000 mPa-s (SiO5000) silicon oil were calculated.

Results

SiO fill between 80% to 90% allowed 55% to 78% retinal contact. The superior retina always kept better contact with SiO, regardless of the fill percentage (P < 0.01)