Clinical and radiographic predictors of cardiovascular implantable electronic device lead failure at the time of initial implantation.


ABSTRACT:

Objective

To assess the clinical and radiographic factors associated with lead failure by comparing subjects with lead failure within 10 years of implantation with an implant-year-matched group without lead failure.

Methods

A case-control study with 49 subjects who received Cardiac Implantable Electronic Device (CIED) between January 1, 1999 and July 31, 2008 and developed lead failure within 10 years of implantation in a single center. The control group consisted of subjects (n = 54) with normally functioning leads matched one-to-one by implant year.

Results

Among the failure group, the meantime from implantation to device lead failure was 4.70 ± 2.94 years. Older age at implantation was associated with a lower likelihood of lead failure (Odds Ratio (OR) = 0.28 (75 v